Problem
of exit block and hence failure to capture with reference to pulse amplitude
and pulse duration
1.
Failure to capture
a)
Occurs when
pacemaker spike is not followed by a P &/or QRS
b)
Technical
causes
i)
loose
connection
ii)
lead
dislodgment
iii)
inadequate
mA output
c)
Increased
pacing threshold
i)
Inflammation
& fibrosis around pacing electrode
ii)
Myocardial
ischaemia/infarction
iii)
Acute
electrolyte/acid-base disturbances
a)
Elevated
potassium
b)
Elevated
pCO2
c)
hypoxia
d)
hyperglycaemia
iv)
Metabolic
disturbance
a)
Hypothermia
v)
Drugs
a)
Antiarrhythmics
b)
Beta-blockade
2.
Treatment
a)
If due to
increased pacing threshold
i)
Increase
output (mA)
ii)
?increase
pulse duration (seldom a problem )
iii)
Often
pacing thresholds are different for both electrodes; use electrode with lowest
threshold
b)
Correct underlying
physiological disturbance
c)
Atrial
capture is not achievable if develop AF or Af