When 2 or more items of equipment are connected to a patient with a connection from one to an invaded point within the body, and especially adjacent to the myocardium, the danger of induced VF exists


1.
Electrical
(shock) hazard is possible when both:
a)
There is a
path by which the electrical current can flow from an electrical device through
a patient to ground
b)
There is
some fault in the electrical grounding of the apparatus
2.
If the
circuit includes an intracardiac monitor such as a CVP line, the leakage
current can be transmitted directly to the heart (microshock hazard)
a)
Any
equipment that bypasses skin resistance (epicardial pacemakers, CVC) should be
handled with care (eg rubber gloves) to avoid any leakage current reaching the
patient
3.
The
addition of more pieces of equipment directly increases the risk of exposure to
electrical hazard due to to the fact that possible exposure to faults within
the individual pieces of equipment increases
4.
Equipotential
grounding
a)
A major
problem in monitoring the patient with cardiac disease is the multiple
connection to electrical devices
b)
To provide
safety through proper grounding, each device should be attached to a common ground wire, will drain all current leakage to a single ground
point
c)
A single ground
reference eliminates current flow caused by slight variances in ground
potential [“Cardiac protected area”]
d)
A panel of
outlets near the patient, connected to a common ground, should be the only
outlets used
e)
Connection
to a remote outlet may result in current flow from one electrical device to
another with a slightly different potential or ground which can pass through
the patient even when the equipment is turned off
5.
Capacitance
leakage
a)
Looping or
kinking of electrical cable should be avoided to minimise resistive or
capacitance leakage