Basic principles of design of an ECG amplifier - sensitivity [potentials]; frequency response; common mode rejection ratio

 

1.                  Amplifier

a)                  Biomedical parameters are difficult to measure due to their relatively low energies and similarity of signals from organ to organ. This similarity makes discrimination of a particular signal difficult. Not only does the transducer have a certain frequency response and sensitivity to detect a particular signal accurately, but the signal generated by the transducer requires processing (eg, impedance matching or filtering of noise) and amplification.

b)                 Monitors require several volts of signal to operate, therefore an amplifier is necessary to boost the size of the biological signal 1000’s times before display (eg R wave usually < 2 mv).

 

2.                  Frequency response

a)                  Amplifier for a particular biological signal must be able to respond to the range of frequencies concerned

b)                 Interference in the signal can be minimised if the bandwidth of the equipment used is selected to match the signal correctly

c)                  Frequency content of the signal & the frequency response of the monitor are important determinants of the accuracy of the signal representation and interpretation

d)                 Range of frequencies:

i)                    EEG: 1—60 Hz

ii)                  ECG: 0.5—100 Hz

iii)                 Notice both include potential for 50 Hz power line interference

iv)                AHA recommend frequency response of 0.05—100 Hz for ECG analysis

e)                  Changes in the ECG waveform that occur:

i)                    Rapidly (QRS) contain higher frequency information

ii)                  More slowly (ST segment) contain lower frequency information

 

 

 

3.                  Common mode rejection ratio

a)                  Ability of an differential amplifier to ignore the interference common to both electrodes

b)                 Biological signals

i)                    Very small signal

ii)                  Requires a very high CMRR

a)                  > 1000:1

iii)                 A high CMRR means that the amplifier strongly attenuates signals that are common to both inputs, but amplifies the difference between the two signals

 

 

4.                  Sensitivities

a)                  The amplifier must be suitable for the voltage range of the signal

b)                 A very sensitive amplifier is needed for EEG measurement as the size of the signal is so small

c)                  Monitors require several volts of signal to operate, therefore an amplifier is necessary to boost the size of the biological signal 1000’s times before display (eg R wave usually < 2 mv)

d)                 The ratio of the voltage at the output of an amplifier to the signal voltage at the input is known as the gain of the amplifier

e)                  ECG: 0.1 — 6 mv

f)                   EEG: 1m — 0.3 mv